Z-2229-2010

SpeedPass Disposable Suture Lariat with Nitinol Wire 25 Degree Left, QTY. 1, Biomet Sports Medicine, Warsaw...

Zimmer GmbH / initiated 2010-06-18 / Terminated / root cause: design

Biomet, Inc. began a recall of SpeedPass Disposable Suture Lariat with Nitinol Wire 25 Degree Left, QTY. 1, Biomet Sports Medicine, Warsaw, IN 46581. Sterile. Part number 904010. Used to... on . The reason FDA records is "The nitinol wire may be fractured or may have the potential to fracture making the suture lariat unusable and possibly delaying surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2229-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
SpeedPass Disposable Suture Lariat with Nitinol Wire 25 Degree Left, QTY. 1, Biomet Sports Medicine, Warsaw, IN 46581. Sterile. Part number 904010. Used to pass suture through tissue.
Product code
MDM Instrument, Manual, Surgical, General Use
Reason for recall
The nitinol wire may be fractured or may have the potential to fracture making the suture lariat unusable and possibly delaying surgery.
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm notified all 29 consignees by An Urgent Medical Device Recall Notice letter on 6/18/2010. The letter identified the affected product and explained the reason for recall. Customers are asked to locate and remove the affected products, follow the instructions on the enclosed "FAX Back Response Form," and fax a copy of the form back prior to return of the affected product. Additionally, if product has been further distributed, then hospital personnel must be notified and given the "Dear Biomet Customer" letter. Questions should be directed to Mary Johnson at 800-348-9500 or 574-372-3983.
Quantity in commerce
301 units total
Distribution
Worldwide Distribution -- USA, including the states of FL, IL, NY, CO, MD, CA, GA, AL, PA, OH, SD, KY, TX, ID, WA, NC, AK, & MN, and Columbia and The Netherlands.
Product codes and lots
Lot numbers: 050260, 863960, and 863970.