Z-2232-2010
Suture Lariat 45 Degree Dog Leg Up, QTY. 1, Biomet Sports Medicine, Warsaw, IN 46581. Sterile. Part number...
Zimmer GmbH / initiated 2010-06-18 / Terminated / root cause: design
Biomet, Inc. began a recall of Suture Lariat 45 Degree Dog Leg Up, QTY. 1, Biomet Sports Medicine, Warsaw, IN 46581. Sterile. Part number 904025. Used to pass suture through tissue on . The reason FDA records is "The nitinol wire may be fractured or may have the potential to fracture making the suture lariat unusable and possibly delaying surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2232-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Suture Lariat 45 Degree Dog Leg Up, QTY. 1, Biomet Sports Medicine, Warsaw, IN 46581. Sterile. Part number 904025. Used to pass suture through tissue.
- Product code
- MDM Instrument, Manual, Surgical, General Use
- Reason for recall
- The nitinol wire may be fractured or may have the potential to fracture making the suture lariat unusable and possibly delaying surgery.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- The firm notified all 29 consignees by An Urgent Medical Device Recall Notice letter on 6/18/2010. The letter identified the affected product and explained the reason for recall. Customers are asked to locate and remove the affected products, follow the instructions on the enclosed "FAX Back Response Form," and fax a copy of the form back prior to return of the affected product. Additionally, if product has been further distributed, then hospital personnel must be notified and given the "Dear Biomet Customer" letter. Questions should be directed to Mary Johnson at 800-348-9500 or 574-372-3983.
- Quantity in commerce
- 301 units total
- Distribution
- Worldwide Distribution -- USA, including the states of FL, IL, NY, CO, MD, CA, GA, AL, PA, OH, SD, KY, TX, ID, WA, NC, AK, & MN, and Columbia and The Netherlands.
- Product codes and lots
- Lot numbers: 864020 and 864030.