Z-2355-2010

Beckman Coulter Access Thyroglobulin Antibody II Reagent, Part Number: A32898. The Access Thyroglobulin...

Siemens Healthcare Diagnostics Inc. / initiated 2010-06-29 / Terminated

Beckman Coulter Inc. began a recall of Beckman Coulter Access Thyroglobulin Antibody II Reagent, Part Number: A32898. The Access Thyroglobulin Antibody II (TgAb) assay intended use is for the... on . The reason FDA records is "The affected Access Thyroglobulin Antibody II reagent may produce inaccurate results due to a change in the thyroglobulin that was used in the kits". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2355-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter Access Thyroglobulin Antibody II Reagent, Part Number: A32898. The Access Thyroglobulin Antibody II (TgAb) assay intended use is for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems.
Product code
JNL Immunochemical, Thyroglobulin Autoantibody
Reason for recall
The affected Access Thyroglobulin Antibody II reagent may produce inaccurate results due to a change in the thyroglobulin that was used in the kits.
FDA root cause
Other
Action
Beckman Coulter sent an "Urgent Product Corrective Action" letter dated June 29, 2010, with attached Customer Response form to the affected customers. The letter identified the affected product and issue. Customers were instructed to: (1) Discontinue use of Access Thyroglobulin Antibody II reagent lot numbers 916271, 917073, and 918433. (2) Review results in the context of the total clinical presentation of the patient, and rerun questionable results using another reagent. (3) Complete and return a response form within 10 days. (4) For replacement contact Beckman Coulter at (US) 1-800-526-3821, (Canada) 1-800-463-7828. Customer Technical Support can be contacted at 1-800-854- 3633.
Quantity in commerce
1071 units
Distribution
Worldwide distribution : United States, including states of CA, CO, FL, GA. IA, IL, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI, and country of Canada.
Product codes and lots
Lot Numbers: 917073, 918433 & 916271