Linvatec Corp. dba ConMed Linvatec began a recall of REF 8570 MENISCAL REPAIR NEEDLE, STERILE, Rx ONLY. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908 USA. Intended for use in meniscal repairs on . The reason FDA records is "This recall was due to the possibility that the products may have a breach in the poly-Tyvek pouch that could potentially compromise the sterility of the contents". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2245-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Linvatec Corporation
- Product
- REF 8570 MENISCAL REPAIR NEEDLE, STERILE, Rx ONLY. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908 USA. Intended for use in meniscal repairs.
- Product code
- HWQ Passer
- Reason for recall
- This recall was due to the possibility that the products may have a breach in the poly-Tyvek pouch that could potentially compromise the sterility of the contents.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Each consignee will be notified of the recall by an Urgent Medical Device Recall Notification letter sent by FedEx or an equivalent method. The letter identified the affected product, explained the reason for recall, and health risk. Consignees were instructed to examine their inventory and segregate the affected product for return. Customers are to complete and return the Reply Form and to notify their customers if the product was further distributed. Consignees are instructed to return any remaining product to ConMed Linvatec. Questions should be directed to the Customer Service department at 800-535-8536.
- Quantity in commerce
- 35,451 total
- Distribution
- Worldwide Distribution -- United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Republic of Korea, Netherlands, Portugal, Serbia, Singapore, Slovenia, El Salvador, Thailand, Taiwan, South Africa, and throughout the U.S.
- Product codes and lots
- Lot Number: 46689, 12055, 73890, 82337, 83411, 83409, 82336, 83408, 82335, 96154, 97820, 96153, 96152, 102390, 106101, 106976, 107301, 117737, and 117736.