St. Jude Medical, Engage TR Introducer, 6 F - ACT (2.25 mm), 25 cm length, .035" Max Guidewire O.D., Rx...
St. Jude Medical Cardiovascular Division began a recall of St. Jude Medical, Engage TR Introducer, 6 F - ACT (2.25 mm), 25 cm length, .035" Max Guidewire O.D., Rx, Sterile EO, REF C408513, 100017549 on . The reason FDA records is "Potential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2180-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- St. Jude Medical, Engage TR Introducer, 6 F - ACT (2.25 mm), 25 cm length, .035" Max Guidewire O.D., Rx, Sterile EO, REF C408513, 100017549.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Potential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak around the Introducer hub and strain relief.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Consignees were sent "Urgent Medical Device Recall Notice" dated June 24, 2010 from St. Jude Medical . The letter was addressed to Cath Lab Manger/ Risk Manager. The letter described the problem and the product involved. Advised consignees to discontinue use of the product and remove from inventory. Sales Representatives were also sent a St. Jude Medical Memorandum "Field Action Notice" dated June 24, 2010 and a reconciliation form. The memorandum described the problem and product. Advised Sales Rep to visit the affected accounts and to performed the assigned tasks. A St. Jude Medical sales representative will be contacting for removal and replacement of the affected product. For additional information, please contact St. Jude Medical at 651-491-4470.
- Quantity in commerce
- 5 US, 10 OUS
- Distribution
- Worldwide Distribution including USA, Italy, Denmark, France, Switzerland, England, Finland, Austria, Spain and Estonia.
- Product codes and lots
- Batch in US and OUS: 3107645