Z-0823-2011
Steris Reliance EPS Endoscope Processing System , manufactured by Corporation STERIS Canada, Quebec, Canada...
STERIS Corporation / initiated 2010-07-02 / Terminated
Steris Corporation began a recall of Steris Reliance EPS Endoscope Processing System , manufactured by Corporation STERIS Canada, Quebec, Canada. Washing and high level disinfection of up to two... on . The reason FDA records is "Reliance EPS had internal water leaks and interrupted cycles when air leaks triggered false Fault 38 alarm that DG/HLD cup has not been replaced". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0823-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- Steris Reliance EPS Endoscope Processing System , manufactured by Corporation STERIS Canada, Quebec, Canada. Washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-resistant, semi-critical endoscope devices.
- Product code
- MLR Sterilizer, Chemical
- Reason for recall
- Reliance EPS had internal water leaks and interrupted cycles when air leaks triggered false Fault 38 alarm that DG/HLD cup has not been replaced.
- FDA root cause
- Other
- Action
- Steris issued an "Urgent Field Correction Notice" dated July 2, 2010 to each affected US consignee, notifying them of the issue and affected device. Steris Field Service Representatives and Dealers will contact each affected customer to schedule a service visit to install system upgrades. In the event of an interrupted cycle prior to the upgrade, consignees were instructed that the load should be handled in accordance with the Reliance EPS Operators Manual. Steris can be contacted at 440-392-7043.
- Quantity in commerce
- 936
- Distribution
- Worldwide distribution: USA, including states of AK, AL, AZ, CA, CO, DE, FL, ID, IL, KS, KY, MD, MI, MN, MO, NC, NV, NY, OH, PA, SC, VA, WA, WI, and WY, and countries of Algeria, Australia, Belarus, Canada, France, Germany, Greece, India, Italy, Japan, Kuwait, Lebanon, New Zealand, Russian Federation, Saudia Arabia, Spain, Turkey, United Arab Emirates and United Kingdom.
- Product codes and lots
- Model #'s: MB30001, MB30002, MB30003, MB000077, MB000078, MB000079, MB000080 & MB000081