medCOMP(R) MRCTI66841 6.6F Plastic Dignity Low Profile CT Port with Pre-Attached Chronoflex(R) Polyurethane...
Medical Components, Inc dba MedComp began a recall of medCOMP(R) MRCTI66841 6.6F Plastic Dignity Low Profile CT Port with Pre-Attached Chronoflex(R) Polyurethane Catheter - 1 Unit. Catalog #MRCTI66841 on . The reason FDA records is "Recalled product was packaged with the incorrect port and does not match the label. The kit is labeled as a 6.6F "Dignity" Low Profile CT Port. The kit contains a 6.6F "Pro-Fuse" Low Profile CT Port". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2235-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medical Components, Inc.
- Product
- medCOMP(R) MRCTI66841 6.6F Plastic Dignity Low Profile CT Port with Pre-Attached Chronoflex(R) Polyurethane Catheter - 1 Unit. Catalog #MRCTI66841.
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- Recalled product was packaged with the incorrect port and does not match the label. The kit is labeled as a 6.6F "Dignity" Low Profile CT Port. The kit contains a 6.6F "Pro-Fuse" Low Profile CT Port.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Medcomp sales reps and foreign distributor were notified via email on 7/1/10. Product is to be returned to Medcomp.
- Quantity in commerce
- 47 units
- Distribution
- Worldwide Distribution -- USA, including states of NC and TX and country of Israel.
- Product codes and lots
- Lot #MAZL630, Exp. 04/2015
FIRST ARTICLE / WEEKLY