Z-0067-2011

SoftPath GUI Releases 3.17 through 4.1.0. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760...

SCC Soft Computer / initiated 2007-03-28 / Terminated / root cause: software

SCC Soft Computer began a recall of SoftPath GUI Releases 3.17 through 4.1.0. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Manufacture/Distribution Dates: Version 3.17.5... on . The reason FDA records is "SCC Soft Computer, Clearwater, FL issued a Correction for SoftPath GUI Releases 3.17 and 4.1 in March 2007 for clients with high volume output of final reports who use the Batch Print for Final Reports in the Scheduler...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0067-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SCC Soft Computer
Product
SoftPath GUI Releases 3.17 through 4.1.0. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Manufacture/Distribution Dates: Version 3.17.5 10/14/2003. Version 3.17.6 03/28/2003. Version 3.17.7 07/02/2003. Version 3.17.8 03/03/2004. Version 4.1.0 07/03/2003. Version 4.1.1 10/03/2003. Intended for use in the Pathology Lab for receiving and accessioning specimens, documenting tissue processing, producing material labels, generating final diagnostic reports, and creating billing and management reports for surgical, GYN, NON-GYN, and Autopsy cases.
Product code
LNX
Reason for recall
SCC Soft Computer, Clearwater, FL issued a Correction for SoftPath GUI Releases 3.17 and 4.1 in March 2007 for clients with high volume output of final reports who use the Batch Print for Final Reports in the Scheduler, there is a potential for some reports not printing. This issue was corrected for the one affected user in October 2003.
Root cause tag
software
FDA root cause
Software design
Action
SCC Soft Computer sent a SAFETY NOTIFICATION TASK VERBIAGE dated January 5, 2004, to one client. The Notice identified the product, the problem, and the action to be taken by the client. SCC recommended the client set the Search Date Range to a small enough range to assure that the number of cases scanned during the job would be less than 10,000. The client was asked to acknowledge receipt of the task and to indicate if the Workflow or functionality was used/not used and Agree to use the alternative working solution. For any questions regarding this recall call (727) 780-0100, ext. 4235.
Quantity in commerce
84
Distribution
Nationwide - USA, including the states of CA, CO, CT, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, PA, SD, TX, WI, and WV
Product codes and lots
Versions 3.17.5, 3.17.6, 3.17.7, 3.17.8, 4.1.0, and 4.1.1.