Z-0069-2011

Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and...

SCC Soft Computer / initiated 2007-10-05 / Terminated / root cause: software

SCC Soft Computer began a recall of Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1. SCC Soft Computer, 5400 Tech Data Drive... on . The reason FDA records is "SCC Soft Computer, Clearwater, FL issued a Correction for Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.0.2.40.1, and 4.0.2.41.1 on 10/05/2007. A software anomaly in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0069-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SCC Soft Computer
Product
Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1. SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. Manufacture Dates: 4.0.2.38.1 06/15/07 4.02.39.2 06/29/07, 4.02.40.1 07/19/07, 4.02.41.1 08/20/07. The Result Reporting Interface is a highly versatile interface designed to electronically convey clinical laboratory orders and results for the SCC applications to other recipient information systems.
Product code
LNX
Reason for recall
SCC Soft Computer, Clearwater, FL issued a Correction for Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.0.2.40.1, and 4.0.2.41.1 on 10/05/2007. A software anomaly in the result reporting interface caused result comments entered after an empty line not to be sent to the HIS system. This included interpretative comments, results, and normal range c
Root cause tag
software
FDA root cause
Software design
Action
SCC Soft Computer sent a RISK-TO-HEALTH NOTIFICATION TASK VERBIAGE dated October 5, 2007, through their Task Management System to all affected clients on October 17, 2007. The notice identified the product, the problem and the action clients should take to correct the problem. Clients were to acknowledge receipt of the task notice and to identify if the Workflow or functionality is not used and if they Agree to use alternative working solutions. Each client was asked to grant permission to load the required software to correct the issue. For any questions regarding this recall call (727) 789-0100.
Quantity in commerce
35
Distribution
Worldwide Distribution - USA, including AL, CO, FL, GA, IN, KS, KY, MA, MO, NC, NH, NJ, NY, OH, PA, TN, and WA and the country of Canada
Product codes and lots
Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1.