Z-0095-2011

Surgical Pattie, 1/4" x 1/4" Non--Sterile (Each Unit of sales is 100 cards) of patties on each card)

Cerenovus, Inc. / initiated 2010-07-16 / Terminated / root cause: manufacturing process

Codman & Shurtleff, Inc. began a recall of Surgical Pattie, 1/4" x 1/4" Non--Sterile (Each Unit of sales is 100 cards) of patties on each card) Product code: 24-5403 on . The reason FDA records is "Surgical Pattie x-ray stripe (radio-opaque marker) may become detached". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0095-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cerenovus, Inc.
Product
Surgical Pattie, 1/4" x 1/4" Non--Sterile (Each Unit of sales is 100 cards) of patties on each card) Product code: 24-5403
Product code
HBA Neurosurgical Paddie
Reason for recall
Surgical Pattie x-ray stripe (radio-opaque marker) may become detached
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Codman notified customers by letter dated 7/16/10 issued July 23 through July 26, 2010 via Fed'x overnight advising users that the x-ray marker on Codman Surgical Patties may become detached. Customers are instructed to return product from affected lots. An error in the original letter was discovered and a corrected letter dated July 28, 2010 issued. If there are any questions, please contact Codman at (800) 225-0460.
Quantity in commerce
229 boxes (100 cards/10 patties each card)
Distribution
Nationwide Foreign: Canada, Singapore, Belgium, Mexico, Madrid, Dubai, Tokyo, Brazil, New Zealand, Egypt, South Africa, Singapore,
Product codes and lots
09A0061 09B0310 09C0708  09C0709  09C0709  09E1139  09F1516 09F1517  09G1830  09G1830  09H2170  09J2622  09H2170  09J2622  09L3080 10B0196 10B0317  10C0442  10B0318  10C0442 Lot numbers:  8L2688