Z-2294-2010

Greatbatch OptiSeal Valved PTFE Peelable Introducer, Model 1000093-001, 1000093-002, 1000093-003...

Greatbatch Medical / initiated 2010-07-02 / Terminated / root cause: packaging

Greatbatch Medical began a recall of Greatbatch OptiSeal Valved PTFE Peelable Introducer, Model 1000093-001, 1000093-002, 1000093-003, 1000093-004, 1000093-005, 1000093-006, 1000093-007... on . The reason FDA records is "The outer sterile barrier (pouch) packaging for the 12 Fr device is not intact. No breaches in the inner sterile barrier (tray/tyvek lid) have been identified. The OptiSeal introducer and accessories sealed in the tray...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2294-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Greatbatch Medical
Product
Greatbatch OptiSeal Valved PTFE Peelable Introducer, Model 1000093-001, 1000093-002, 1000093-003, 1000093-004, 1000093-005, 1000093-006, 1000093-007. Sterilized using Ethylene Oxide. Greatbatch Medical, 2300 Berkshire Lane North, Minneapolis, MN 55441 USA. The OptiSeal Valved PTFE Peelable Introducer is designed to reduce blood loss and air intake by providing hemostatic sealing at venous pressures. The device is offered in two configurations, one without a sideport and one with a sideport attached to a segment of extension tubing terminating in a 3-way stopcock.
Product code
DYB Introducer, Catheter
Reason for recall
The outer sterile barrier (pouch) packaging for the 12 Fr device is not intact. No breaches in the inner sterile barrier (tray/tyvek lid) have been identified. The OptiSeal introducer and accessories sealed in the tray are sterile. Infection may result if a sterile field is contaminated by the outer surface of the inner tray.
Root cause tag
packaging
FDA root cause
Packaging
Action
Greatbatch Medical issued an "Urgent Medical Device Recall letter dated July 12, 2010 via email to consignees. The letter described the problem and affected device. Consignees were instructed to immediately discontinue distribution of the product and return the unused units to Greatbatch Medical and to notify their customers to immediately return unused units. For shipping assistance, questions or assistance in notifying accounts, contact Shannon Springer at 763 951-8244 or Kimberly Briggs, 716 759-5702.
Quantity in commerce
211 5-pack boxes
Distribution
Worldwide Distribution: United States, including the states of MN and CA.
Product codes and lots
Model / lot number: 1000093-001 / W1515452; 1000093-002 / W1545383; 1000093-003 / W1519579, W1524143, W1515454; 1000093-004 / W1524146, W1519578, W1520649, W1515441; 1000093-005 / W1515455; 1000093-006 / W1515456; 1000093-007 / W1560536.