Z-2242-2010

CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded...

CIVCO Medical Instruments Co., Inc. / initiated 2010-07-20 / Terminated / root cause: manufacturing process

Civco Medical Instruments Inc began a recall of CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded cover with attached (6cm) NeoGuard tip. The... on . The reason FDA records is "Polyethylene material did not have proper treatment applied during manufacturing. Untreated polyethylene material does not maintain a secure seal with the adhesive tape used to secure the tip to the cover". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2242-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CIVCO Medical Instruments Co., Inc.
Product
CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded cover with attached (6cm) NeoGuard tip. The latex-free surgi tipped covers are utilized to cover ultrasound transducers and prevent cross-contamination from occurring with equipment and patients.
Product code
ITX Transducer, Ultrasonic, Diagnostic
Reason for recall
Polyethylene material did not have proper treatment applied during manufacturing. Untreated polyethylene material does not maintain a secure seal with the adhesive tape used to secure the tip to the cover.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
CIVCO sent a "Urgent - Medical Device Correctdion" letter date July 19, 2010 to consignees describing the reason for recall and requesting they dispose of the affect lot number of product. A response form was enclosed for completion and return to the firm, indicating the amount of product in inventory and requesting a Return Authorization Number. The form was to be returned via FAX or e-mail. Customers may contact CIVO at 800-445-6741.
Quantity in commerce
22 boxes
Distribution
Nationwide distribution: the states of OH, MD, NE, VA, IA, MI, WA, and WI.
Product codes and lots
Lot M538160