CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded...
Civco Medical Instruments Inc began a recall of CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded cover with attached (6cm) NeoGuard tip. The... on . The reason FDA records is "Polyethylene material did not have proper treatment applied during manufacturing. Untreated polyethylene material does not maintain a secure seal with the adhesive tape used to secure the tip to the cover". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2242-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CIVCO Medical Instruments Co., Inc.
- Product
- CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded cover with attached (6cm) NeoGuard tip. The latex-free surgi tipped covers are utilized to cover ultrasound transducers and prevent cross-contamination from occurring with equipment and patients.
- Product code
- ITX Transducer, Ultrasonic, Diagnostic
- Reason for recall
- Polyethylene material did not have proper treatment applied during manufacturing. Untreated polyethylene material does not maintain a secure seal with the adhesive tape used to secure the tip to the cover.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- CIVCO sent a "Urgent - Medical Device Correctdion" letter date July 19, 2010 to consignees describing the reason for recall and requesting they dispose of the affect lot number of product. A response form was enclosed for completion and return to the firm, indicating the amount of product in inventory and requesting a Return Authorization Number. The form was to be returned via FAX or e-mail. Customers may contact CIVO at 800-445-6741.
- Quantity in commerce
- 22 boxes
- Distribution
- Nationwide distribution: the states of OH, MD, NE, VA, IA, MI, WA, and WI.
- Product codes and lots
- Lot M538160