Z-0012-2011

Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, including...

Philips North America LLC / initiated 2008-08-15 / Terminated

Philips Medical Systems North America Co. Phillips began a recall of Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, including specialist areas like trauma or pediatric work... on . The reason FDA records is "When a patient is selected in the patient list and examination tab pressed with background query running at the same time, the wrong patient may be selected". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0012-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, including specialist areas like trauma or pediatric work, excluding mammography. It is designed for radiographic examination of the standing or seated patient or the recumbent patient in combination with a mobile X-ray table (trolley). The intended use is from direct digital imaging using the built in flat panel detector and in addition for free exposures on radiographic cassettes.
Product code
KPR System, X-Ray, Stationary
Reason for recall
When a patient is selected in the patient list and examination tab pressed with background query running at the same time, the wrong patient may be selected.
FDA root cause
Other
Action
Philip Medical System sent an "URGENT-Device Correction" letter dated August 13, 2008 to their customers. The letter describes the recalled product as Essenta DR Patient and Image Data Software (SW) release 1.0. The firm informs the customers that when patient is selected in patient list and examination tab pressed with background query running at the same time, there is a risk that the wrong patient is selected on the Essenta DR Computed Radiography System. Customers are advised to verify the selected patient's information before releasing an exposure while the firm works on upgrading the Software. All affected units will receive a software upgrade to correct the issue. Customers should contact Philips Customer Care Service at 800-772-9377, #5, #4, #1 and reference FCO 712000031 for any questions or support concerning the recall.
Quantity in commerce
23 units
Distribution
Nationwide distribution: CA, GA, FL, KS, MN, NY, OH, OR, SC, TN, TX, UT, and WA.
Product codes and lots
Devices were identified as SITE numbers:  553491, 554498, 554944, 553270, 554358, 556089, 556275, 555457, 556080, 557203, 41446107, 41446304, 41443692, 41446097, 41445359, 41445529, 41585399, 41586528, 41796147, 41958550, 42252302, 42616819, and 42865052.