Z-2351-2010
CELL-DYN 3700 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste...
Abbott Medical / initiated 2010-04-16 / Terminated / root cause: manufacturing process
Abbott Laboratories began a recall of CELL-DYN 3700 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802 on . The reason FDA records is "Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2351-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- CELL-DYN 3700 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Product Recall letters were sent on 04/16/2010. The letters identified the affected product, explained the reason for recall, the patient impact/user safety, stated the necessary actions, and contact information. Users are to discontinue use until the Waste Chamber(s) have been replaced and follow the instructions in the Hazards section. In addition, appropriate protective equipment needs to be worn along with following their recommended laboratory procedures dealing with spills. Users are to contact their local Abbott Customer Support rep. US customers should be direct their questions to Customer Support.
- Quantity in commerce
- 69 units total
- Distribution
- Worldwide Distribution -- CA, CT, ID, IN, and MA; and, countries of Australia, Brazil, Canada, China, Colombia, Mexico, Puerto Rico, Singapore, and Uruguay.
- Product codes and lots
- Reservoir Date Codes Dec 17 2009 to March 17, 2010. Part Number/Serial numbers-- CS: 02H30-01/20478AN, 20479AN SL: 02H31-01/22470AK, 21690AK, 22062AK, 21699AK, 20756AK, 22459AK, 21302AK, 21601AK, 22753AK