Z-2352-2010

CELL-DYN Ruby, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir...

Abbott Medical / initiated 2010-04-16 / Terminated / root cause: manufacturing process

Abbott Laboratories began a recall of CELL-DYN Ruby, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802 on . The reason FDA records is "Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2352-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
CELL-DYN Ruby, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802.
Product code
GKZ Counter, Differential Cell
Reason for recall
Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Product Recall letters were sent on 04/16/2010. The letters identified the affected product, explained the reason for recall, the patient impact/user safety, stated the necessary actions, and contact information. Users are to discontinue use until the Waste Chamber(s) have been replaced and follow the instructions in the Hazards section. In addition, appropriate protective equipment needs to be worn along with following their recommended laboratory procedures dealing with spills. Users are to contact their local Abbott Customer Support rep. US customers should be direct their questions to Customer Support.
Quantity in commerce
69 units total
Distribution
Worldwide Distribution -- CA, CT, ID, IN, and MA; and, countries of Australia, Brazil, Canada, China, Colombia, Mexico, Puerto Rico, Singapore, and Uruguay.
Product codes and lots
Reservoir Date Codes Dec 17 2009 to March 17, 2010. Part Number/Serial numbers-- 08H67-01/35286BG-35318BG, 34891BG, 34641BG, 34408BG, 34890BG, 34021BG