NeoMarkers Rabbit Monoclonal anti-Human Cyclin D1 Antibody (Clone SP4), immunochemistry Reagent and Kit, For...
Lab Vision Corporation began a recall of NeoMarkers Rabbit Monoclonal anti-Human Cyclin D1 Antibody (Clone SP4), immunochemistry Reagent and Kit, For In Vitro Diagnostic use, manufactured by Lab... on . The reason FDA records is "Two lots may be contaminated with Ki-67 antibody, which could cause abnormal staining patterns, and produce false positive results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2424-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Lab Vision Corp.
- Product
- NeoMarkers Rabbit Monoclonal anti-Human Cyclin D1 Antibody (Clone SP4), immunochemistry Reagent and Kit, For In Vitro Diagnostic use, manufactured by Lab Vision Products, Fremont, CA, a subsidiary of ThermoFisher Scientific
- Product code
- NJT Immunohistochemistry Reagents And Kits
- Reason for recall
- Two lots may be contaminated with Ki-67 antibody, which could cause abnormal staining patterns, and produce false positive results.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Thermo Fisher Scientific issued an "Advisory Notice" letter dated March 5, 2010 to all domestic customers and international distributors. The letter described the product, problem and action to be taken. Customers were instructed to: 1) discard the affected lots, and 2) retest specimens using a replacement lot supplied by the firm, and 3) complete and return the Confirmation of Receipt Form via fax to 1-510-979-5489 International distributors were instructed to: 1) place the affected lots and replace with a replacement lot supplied by the firm, and 2) provide customers with copies of the customer letter, and 3) complete and return the Confirmation of Receipt Form via fax to 1-510-979-5489 For questions, contact Thermo Fisher Scientific at 1-800-828-1628 or 1-510-771-1595.
- Quantity in commerce
- 103 units
- Distribution
- Worldwide Distribution: USA, including the states of CA, SC, IL NY, WI, PA, MN, AL, WA, MA, and KS, and the coutnries of Austria, Australia, Denmark, Finland, France, Germany, Hong Kong, Italy, Japan, Malaysia, the Netherlands, Norway, Pakistan, Poland, Portugal, Russia, the UK, and Turkey.
- Product codes and lots
- lots 9104S911A through 9104S911E and 9104S1001A, Catalog number RM-910, Expiration dates November 2011 and January 2012.