Z-0287-2011

enpath medical Myopore Sutureless Myocardial Pacing Lead, Bipolar (labeled as Unipolar) Connector, , REF...

Greatbatch Medical / initiated 2010-05-13 / Terminated / root cause: manufacturing process

Greatbatch Medical began a recall of enpath medical Myopore Sutureless Myocardial Pacing Lead, Bipolar (labeled as Unipolar) Connector, , REF 511212. 54 cm. Distributed by Guidant/CPI. Contents in... on . The reason FDA records is "The Myopore Bipolar Epicardial Leads packaging is labeled as unipolar leads. The lead body itself is correctly labeled as bipolar. This could result in a bipolar lead being implanted with a pulse generator operating in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0287-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Greatbatch Medical
Product
enpath medical Myopore Sutureless Myocardial Pacing Lead, Bipolar (labeled as Unipolar) Connector, , REF 511212. 54 cm. Distributed by Guidant/CPI. Contents in unopened, undamaged package have been sterilized by ethylene oxide gas. Manufactured by enpath medical, Inc., 2300 Berkshire Lane North, Minneapolis, MN 55441 USA.
Product code
DTB Permanent Pacemaker Electrode
Reason for recall
The Myopore Bipolar Epicardial Leads packaging is labeled as unipolar leads. The lead body itself is correctly labeled as bipolar. This could result in a bipolar lead being implanted with a pulse generator operating in unipolar mode.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Greatbatch Medical issued an Urgent Medical Device Recall letter dated May 13, 2010 to their distributor. The letter described the problem and product involved, and gave the action required which included: 1) immediately discontinue distribution of the units and return any unused units to Greatbatch Medical. 2) Reconcile medical recordsbetween BSC and Greatbatch Medical to confirm implant data 3) Confirm pulse generator type (bipoloar, unipolar, or unknown) used with each serial number listed Greatbatch will work with customers to contact physicians as necessary regarding patient monitoring. Greatbatch Medical can be contacted at 763-951-8312.
Quantity in commerce
34
Distribution
Worldwide Distribution: to USA and Europe, through distributor in MN.
Product codes and lots
serial numbers: 123070, 123072, 123073, 123075 thru 123092, 123094, 123095, 123097 thru 123104, 123431, 123432, 123433.