Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2316-2010

Boston Scientific Ultraflex Uncovered Esophageal Stent System - Distal Release (18/23/15) Material/Catalog...

Boston Scientific Corporation / initiated 2010-07-15 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific Ultraflex Uncovered Esophageal Stent System - Distal Release (18/23/15) Material/Catalog Number: M00513720 on . The reason FDA records is "Units have excessive fraying of the deployment suture thread, as a result, these devices may experience a suture break during deployment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2316-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Ultraflex Uncovered Esophageal Stent System - Distal Release (18/23/15) Material/Catalog Number: M00513720.
Product code
ESW Prosthesis, Esophageal
Reason for recall
Units have excessive fraying of the deployment suture thread, as a result, these devices may experience a suture break during deployment.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Boston Scientific issued "Urgent Medical Device Recall - Immediate Action Required" letters dated July 15,2010 to consignees. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory. Foreign accounts received notification translated into the local language and with changes that are applicable to local regulations. Boston Scientific can be contacted at 1-508-683-4427.
Quantity in commerce
0
Distribution
Worldwide Distribution: USA, and the countries of: Austria, Germany, Finland, France, UK, Ireland, Iran, Italy, the Netherlands, Portugual, Spain, and Sweden
Product codes and lots
Lot/Batch Number: 13450513