Boston Scientific Ultraflex Covered Esophageal Stent System - Proximal Release (18/23/10/7)...
Boston Scientific Corporation began a recall of Boston Scientific Ultraflex Covered Esophageal Stent System - Proximal Release (18/23/10/7) Material/UPN/Catalog Number: M00513840 on . The reason FDA records is "Units have excessive fraying of the deployment suture thread, as a result, these devices may experience a suture break during deployment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2319-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Ultraflex Covered Esophageal Stent System - Proximal Release (18/23/10/7) Material/UPN/Catalog Number: M00513840
- Product code
- ESW Prosthesis, Esophageal
- Reason for recall
- Units have excessive fraying of the deployment suture thread, as a result, these devices may experience a suture break during deployment.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Boston Scientific issued "Urgent Medical Device Recall - Immediate Action Required" letters dated July 15,2010 to consignees. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory. Foreign accounts received notification translated into the local language and with changes that are applicable to local regulations. Boston Scientific can be contacted at 1-508-683-4427.
- Quantity in commerce
- 10 units
- Distribution
- Worldwide Distribution: USA, and the countries of: Austria, Germany, Finland, France, UK, Ireland, Iran, Italy, the Netherlands, Portugual, Spain, and Sweden
- Product codes and lots
- Lot/Batch Number: 13441743