Siemens brand ARTISTE MV, Artiste Digital Linear Accelerators in combination with the Flat Panel, Part No...
Siemens Medical Solutions USA, Inc began a recall of Siemens brand ARTISTE MV, Artiste Digital Linear Accelerators in combination with the Flat Panel, Part No. 8139789, To deliver x-ray radiation for therapeutic... on . The reason FDA records is "The robotic lift armature shaft may break and allow the Flat Panel Positioner with the digital flat panel imager to fall into the Flat Panel cradle. If the positioner is over a patient, there is a potential for injury". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0449-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens brand ARTISTE MV, Artiste Digital Linear Accelerators in combination with the Flat Panel, Part No. 8139789, To deliver x-ray radiation for therapeutic treatment of cancer. Product is manufactured and distributed by Siemens Medical Solutions, Concord, CA
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- The robotic lift armature shaft may break and allow the Flat Panel Positioner with the digital flat panel imager to fall into the Flat Panel cradle. If the positioner is over a patient, there is a potential for injury.
- FDA root cause
- Other
- Action
- Siemens sent an Urgent Medical Device Correction letter to all on January 27, 2010, via phone, Fax, certified mail and/or delivery via Siemens Customer Service Group to all its direct consignees, informing them of the affected products, the problem, and provided instructions on the recall. Consigneees were to be contacted by an ARTISTE Service Representative for an appointment for corrective action. Consignees were instructed to include Customer Safety Advisory Notice in their ARTISTE System Owner Manual. For questions regarding this recall call (925) 246-8407.
- Quantity in commerce
- 106
- Distribution
- Worldwide Distribution - USA including CA, DE, FL, LA, MA, MI, MO, NC, NE, NJ, OH, PA, TN, UT, WI, AND WV and Egypt, Germany, Japan, Lebanon, Malaysia, New Zealand, Poland, Russia, Spain, Sweden, and Turkey
- Product codes and lots
- All units--Dates of manufacture: July 17, 2008 to September 19, 2008.