Z-0106-2011

Varian brand TrueBeam and TrueBeam STx, Model Number:TMX- H19, Distributed by and Manufactured by: Varian...

Varian Medical Systems, Inc. / initiated 2010-05-24 / Terminated / root cause: software

Varian Medical Systems Oncology Systems began a recall of Varian brand TrueBeam and TrueBeam STx, Model Number:TMX- H19, Distributed by and Manufactured by: Varian Medical Systems Inc., Palo Alto, CA on . The reason FDA records is "Due to a software anomaly, guidance-based couch shift values may not be applied as expected when the operator selects "Apply Shift" and presses Motion Enable buttons. This can result in the actual patient position...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0106-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Varian brand TrueBeam and TrueBeam STx, Model Number:TMX- H19, Distributed by and Manufactured by: Varian Medical Systems Inc., Palo Alto, CA
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Due to a software anomaly, guidance-based couch shift values may not be applied as expected when the operator selects "Apply Shift" and presses Motion Enable buttons. This can result in the actual patient position differing from that position indicated by the imaging application affecting poistion accuracy and precision.
Root cause tag
software
FDA root cause
Software design
Action
Varian Medical Systems issued an Urgent Medical Device Correction letter date dated May 24, 2010 to users identifying the affected device, the issue prompting the correction, and the actions to be taken by the user. Customers were asked to set couch tolerance values to small numbers, and for radiosurgical treatment tolerance values to be set to zero. A Varian representative will contact customers to schedule installation of a software version correcting the issue. US Customers can contact Varian at 1-888-827-4265,. European customers can contact Varian at +41 41 749 8844.
Distribution
Worldwide Distribution: Two consignees, one in the US and one in Switzerland.
Product codes and lots
H191001, H191003