Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332. Intended to be used to contain medical devices...
Advanced Sterilization Products began a recall of Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332. Intended to be used to contain medical devices that are being terminally sterilized in STERRAD... on . The reason FDA records is "The recall was initiated after Advanced Sterilization Products (ASP) discovered a non-conforming lot. The misaligned perforations may have caused the seal to be incomplete on some of the pouches, creating a potential...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1393-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Advanced Sterilization Products
- Product
- Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332. Intended to be used to contain medical devices that are being terminally sterilized in STERRAD Sterilization Systems.
- Product code
- KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
- Reason for recall
- The recall was initiated after Advanced Sterilization Products (ASP) discovered a non-conforming lot. The misaligned perforations may have caused the seal to be incomplete on some of the pouches, creating a potential breach for pouches being used to store sterilized equipment on shelves.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Advanced Sterilization Products (ASP) released Urgent: Product Recall notifications to its consignees identifying the affected product and explaining the reason for recall. The letter also discussed the impact of using this affected product and return instructions. The firm asked customers to examine their inventory, discontinue use of Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332 and to return all unused affected products to Stericycle. Customers are to complete the Business Reply Card and return it to Stericycle.
- Quantity in commerce
- 470
- Distribution
- Worldwide Distribution -- USA, Argentina, Australia, Canada, Chile, Czech Republic, Hong Kong, India, Japan, Panama, Peru and Puerto Rico.
- Product codes and lots
- Lot Number: 2020
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
