Z-2201-2011

Software CD is labeled in part: "BLOOD VIRUS DIVISION***BIO-RAD EVOLIS System APF Version 4.4 Software...

Bio-Rad Laboratories Inc. / initiated 2010-07-19 / Terminated / root cause: software

Bio-Rad Laboratories began a recall of Software CD is labeled in part: "BLOOD VIRUS DIVISION***BIO-RAD EVOLIS System APF Version 4.4 Software Version 2.00 2009/05***" The EVOLIS Assay Protocol Files... on . The reason FDA records is "The EVOLIS Assay Protocol Files Software is configured in such a way that may result in a miscalculation of quantitative test results and/or an error in reporting positive/negative sample values". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2201-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
Software CD is labeled in part: "BLOOD VIRUS DIVISION***BIO-RAD EVOLIS System APF Version 4.4 Software Version 2.00 2009/05***" The EVOLIS Assay Protocol Files (APF) Software contains assay-specific instructions necessary for the EVOLIS Microplate processing system to process enzyme immunoassays on the EVOLIS instrument, perform data reduction and generate reports.
Product code
LFX Enzyme Linked Immunoabsorbent Assay, Rubella
Reason for recall
The EVOLIS Assay Protocol Files Software is configured in such a way that may result in a miscalculation of quantitative test results and/or an error in reporting positive/negative sample values.
Root cause tag
software
FDA root cause
Software design
Action
Bio-Rad Laboratories sent an Urgent Product Correction letter dated July 19, 2010, to all customer sites via FedEx. There are two issues related to The EVOLIS Assay Protocol Files (APF) Software that could potentially affect patient results. Issue 1 involves quantitative sample test/calculations/results interpretation for assays with linear regression, point to point or 4PL data reduction. Issue 2 involves sample test results with " Result Report Flags''. A software patch to correct these two issues will be sent to each customer site. Bio-Rad Laboratories will work with each customer to complete documentation confirming that the software patch has been successfully installed on each instrument. If you have any questions or concerns regarding these two issues, please contact Technical Support at 1-800-224-6723, option 2 and then option 3.
Quantity in commerce
272
Distribution
Nationwide Distribution -- Including Puerto Rico.
Product codes and lots
APF Version 4.4 Software Version 2.00 2009/05 LOT=V4.4-0505-2