Z-0930-2011

i-STAT G3+ Cartridge; pH, pCO2, PO2; IVD Abbott Point of Care The test for PCO2, as part of the i-STAT...

Abbott Medical / initiated 2010-08-03 / Terminated

Abbott Point of Care Inc. began a recall of i-STAT G3+ Cartridge; pH, pCO2, PO2; IVD Abbott Point of Care The test for PCO2, as part of the i-STAT System, is intended for use in the vitro quantification... on . The reason FDA records is "Two cartridge lots were identified that may generate falsely elevated PCO2 results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0930-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
i-STAT G3+ Cartridge; pH, pCO2, PO2; IVD Abbott Point of Care The test for PCO2, as part of the i-STAT System, is intended for use in the vitro quantification of carbon dioxide partial pressure in arterial, venous or capillary whole blood. PCO2 along with pH is used to assess acid-base balance.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
Two cartridge lots were identified that may generate falsely elevated PCO2 results.
FDA root cause
Other
Action
Abbott Point of Care Inc. sent an Urgent Recall Notice letter dated July 2010 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the cartridges from the identified lots and follow the instructions on the enclosed reply card to receive replacement cartridges. If cartridges were distributed from the lot numbers to another institution, a copy of the recall letter should be provided to them. For any questions regarding this recall contact Abbott-Point of Care Technical Support at 800-366-8020, Option 1.
Quantity in commerce
55350 total cartridges
Distribution
Worldwide Distribution - USA including AL, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY and the countries of Oman and Turkey.
Product codes and lots
06F01-01, 06F03-02; Lot number: P10033