Z-0004-2011

AortaScan AMI 9700 The AortaScan AMI 9700 is an ultrasound device that projects ultrasound energy into the...

Verathon Inc. / initiated 2010-08-23 / Terminated / root cause: software

Verathon, Inc. began a recall of AortaScan AMI 9700 The AortaScan AMI 9700 is an ultrasound device that projects ultrasound energy into the mid-abdomen to obtain an image of the abdominal... on . The reason FDA records is "The AortaScan AMI 9700 devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0004-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Verathon Inc.
Product
AortaScan AMI 9700 The AortaScan AMI 9700 is an ultrasound device that projects ultrasound energy into the mid-abdomen to obtain an image of the abdominal aorta for aortic diameter measurements
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
The AortaScan AMI 9700 devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.
Root cause tag
software
FDA root cause
Software design
Action
Verathon sent an URGENT Medical Device Recall Notification letter dated August 12, 2010 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Verathon provided an addendum to the User Manual with instructions on how to reset the instrument. Customers were advised to: " Add the enclosed addendum to their current Users Manual. " Update the application software of the device to v2.7.0 or higher. If the customers had questions or needed support for the software update process, they were to contact Verathon Medical Customer Care at (425) 867-1348 or (800) 331-2313 or [email protected].
Quantity in commerce
73 units
Distribution
Worldwide Distribution - USA and the countries of Australia, Canada, China, Colombia, European Union, United Kingdom, Saudi Arabia, and Singapore.
Product codes and lots
Serial numbers:   A7001001, A7001005, A7001040, A7001041, A7001042, A7001043, A7001044, A7001045, A7001046, A7001047, A7001048, A7001049, A7001050, A7001051, A7001052, A7001053, A7001054, A7001055, A7001056, A7001057, A7001062, A7001065, A7001066, A7001067, A7001069, A7001071, A7001089, A7001090, A7001091, A7001092, A7001093, A7001094, A7001095, A7001096, A7001097, A7001098, A7001099, A7001100, A7001101, A7001102, A7001103, A7001106, A7001107, A7001108, A7001109, A7001110, A7001111, A7001112, A7001113, A7001114, A7001115, A7001116, A7001117, A7001118, A7001119, A7001120, A7001121, A7001122, A7001123, A7001124, A7001125, A7001127, A7001128, A7001129, A7001130, A7001131, A7001132, A7001133, A7001134, A7001135, A7001136, A7001138, and A7001139.