Sonosite, Inc. began a recall of The software is ARM version 60.80.101.025 (referred to as Nano 1.0.5). The software is used with the SonoSite NanoMaxx ultrasound system on . The reason FDA records is "Software ARM version 60.80.101.025 (referred to as Nano1.0.5) on the NanoMaxx ultrasound system results in errors when the "print" command is selected. As a result, the displayed MI or TI values may be incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0478-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM SonoSite,Inc.
- Product
- The software is ARM version 60.80.101.025 (referred to as Nano 1.0.5). The software is used with the SonoSite NanoMaxx ultrasound system.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Software ARM version 60.80.101.025 (referred to as Nano1.0.5) on the NanoMaxx ultrasound system results in errors when the "print" command is selected. As a result, the displayed MI or TI values may be incorrect.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Sonosite issued Urgent Medical Device Correction letters dated August 9, 2010 to consignees. Consignees were advised of the software problem and given directions to verify the software version they are using. They are advised to contact SonoSite at 877-657-8118 to obtain a free software upgrade.
- Quantity in commerce
- 69
- Distribution
- Worldwide Distribution: USA and to the countries of Australia, Great Britain, India, Israel, the Netherlands, New Zealand, Norway, and Spain.
- Product codes and lots
- Part code: P11111-35. Serial numbers: 03HQF4, 03HQF3, 03HQF6, 03HQCM, 03HPZK, 03HQ9X, 03H9C3, 03HKN6, 03HPQ9, 03H87H, 03HQ9M, 03HQ9Q, 03HQB2, 03HQB3, 03HQBB, 03HQBX, 03HQBY, 03HQBZ, 03HQC2, 03HQC3, 03HQC4, 03HQC5, 03HQC6, 03HQCC, 03HQCD, 03HQCF, 03HQCG, 03HQCH, 03HQCJ, 03HQCK, 03HQCN, 03HQCP, 03HQCX, 03HQD2, 03HQD3, 03HQD4, 03HQD5, 03HQD8, 03HQD9, 03HQDB, 03HQDD, 03HQDF, 03HQDG, 03HQDH, 03HQDK, 03HQDL, 03HQDN, 03HQDP, 03HQFL, 03HQ77, 03H16Q, 03HKM7, 03HQBH, 03HQD6, 03HQCL, 03HQ6M, 03HQFG, 03HQF5, 03HQB9, 03HQDC, 03HQDT, 03HQDX, 03HQF0, 03HQF2, 03HQF7, 03HPL9, 03HQD7, 03HQC0, and 03HQFJ.