Z-2635-2010

Enseal TRIO Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05

Ethicon Endo-Surgery Inc / initiated 2010-07-09 / Terminated / root cause: manufacturing process

Ethicon Endo-Surgery Inc began a recall of Enseal TRIO Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05 on . The reason FDA records is "The weld that connects the electrode to the active rod was separated. This connection is critical to the proper function of these devices. If the weld is separated, the product will not deliver energy to the jaws...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2635-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon Endo-Surgery Inc
Product
Enseal TRIO Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05
Product code
NLM Laparoscope, General & Plastic Surgery, Reprocessed
Reason for recall
The weld that connects the electrode to the active rod was separated. This connection is critical to the proper function of these devices. If the weld is separated, the product will not deliver energy to the jaws, resulting in the potential for a cut/transection without a seal. Ethicon Endo-Surgery identified this issue during a routine finished goods quality inspection.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Ethicon Endo-Surgery issued an Urgent: Device Recall, Event 2193 letter dated July 19, 2010 identifying the affected device, the issue prompting the recall, and actions to be taken by consignees. Consignees were instructed: Do not use the affected product. Identify and quarantine affected product from inventory. Complete the Business Reply form and fax to Stericycle at 877-597-9580 within three days To return product, place it in a box with a copy of the Business Reply form and mail to Stericycle using the pre-printed UPS shipping label. Replacement product should be received by the customer within 60 days of return of affected product. Customers can contact Stericycle about this recall at 877-243-7314 and referencing Event 2193.
Quantity in commerce
396 instruments
Distribution
Nationwide Distribution to the following states: AZ, AL, AZ, CA, CO, CT, FL, GA, IN, KY, MA, MD, MI, MO, MS, NC, NV, NY, OH, PA, TN, TX, WA, and WV.
Product codes and lots
Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05