Z-2637-2010

Integra CUSA Ultrasonics NXT Ultra Surgical Aspirator System CUSANXT1:Latex Free, Rx only: Integra...

Integra LifeSciences Corporation / initiated 2010-07-29 / Terminated / root cause: design

Integra LifeSciences Corp. began a recall of Integra CUSA Ultrasonics NXT Ultra Surgical Aspirator System CUSANXT1:Latex Free, Rx only: Integra LifeSciences (Ireland) Limited, County Offaly, Ireland... on . The reason FDA records is "A ground fault malfunction at the rectifier retaining screw may cause the aspiration pump in the CUSA NXT Service module to run at maximum speed, and continue to run until the entire system is turned off. This may...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2637-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra CUSA Ultrasonics NXT Ultra Surgical Aspirator System CUSANXT1:Latex Free, Rx only: Integra LifeSciences (Ireland) Limited, County Offaly, Ireland; Distributed by Integra Life Sciences Corporation 311 Enterprise Drive Plainsboro, NJ 08536.
Product code
LFL Instrument, Ultrasonic Surgical
Reason for recall
A ground fault malfunction at the rectifier retaining screw may cause the aspiration pump in the CUSA NXT Service module to run at maximum speed, and continue to run until the entire system is turned off. This may result in excess aspiration that requires medical intervention.
Root cause tag
design
FDA root cause
Device Design
Action
Integra issued Urgent: Product Recall Notification letters dated July 29, 2010 to all customers on July 29, 2010 identifying the affected products, issue prompting the recall, and actions to be taken by the customer. Letters were also forwarded to all Integra Account Reps. Customers were advised to discontinue use of and quarantine the affected service modules. The Surgical Aspirator may be operated without the Service Module, utilizing the hospitals wall aspiration. Company representatives will contact customers to arrange product return and repair of the affected service modules. Customers are to complete and return an enclosed Recall Acknowledgement and Return Form. The firm can be contacted at 609 936-2485.
Quantity in commerce
26 in US, 8 in Europe and 1 in Asia (both unit types.
Distribution
Worldwide Distribution: USA , including the states of AZ, CA, CO, FL, GA, IA, KY, MN, MO, NV, NY, OH, OR, RI, TX, VA, WI, and the countries of Germany, Hungary, Indonesia, Italy, Netherlands, and Switzerland.
Product codes and lots
Serial numbers: NXT1001-1079, NLB0901901IE, NBL0902101IE, NBL0902501IE, NBA1000501IE, NBC1001501IE, NBC1002001IE, NBC1002301IE, NBC1002901IE, NBE1001301IE, NCE1001901IE, NBF1000701IE, NBD1000101IE.