Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number...
Hospira Inc began a recall of Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82... on . The reason FDA records is "The clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2390-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA
- Product code
- FRN Pump, Infusion
- Reason for recall
- The clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Hospira issued an "Urgent: Device Field Correction" Letter dated February 22, 2010 to all consignees, identifying the reason for the recall and affected product. Customers were informed that they will be contacted by a Hospira representative to replace the affected pole clamps. Consignees can contact Hospira Technical Support Operations at 1-800-241-4002.
- Distribution
- Worldwide Distribution: USA (all states and Puerto Rico) and the countries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Singapore. and the UAE.
- Product codes and lots
- All units of these List Numbers.