Z-2354-2010

Titan 3 5K Low-Speed Motor - Star Titan 3, 5,000 rpm- Lube Free

Dentalez, Inc., Stardental Division / initiated 2010-08-11 / Terminated / root cause: labeling and UDI

Dental EZ Stardental Division began a recall of Titan 3 5K Low-Speed Motor - Star Titan 3, 5,000 rpm- Lube Free on . The reason FDA records is "The housing of 5K motor is labeled as a 20K motor". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2354-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Dentalez, Inc., Stardental Division
Product
Titan 3 5K Low-Speed Motor - Star Titan 3, 5,000 rpm- Lube Free
Product code
EFA Handpiece, Belt And/Or Gear Driven, Dental
Reason for recall
The housing of 5K motor is labeled as a 20K motor.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
DentalEZ group/Star Dental Division sent a "Medical Device Recall" letter to consignees on 8/11/10, informing them that the device's 5K motor was mislabeled as a 20K motor. Customers are to stop using of the device, and return to the manufacturer or to the distributor using supplied shipping labels. Replacement devices will be sent to customers at no charge. Consignees are to complete and return a form indicating whether they have any affected devices, and whether customers have been contacted about this recall. Customers can contact DentalEZ at 866-383-4636 .
Quantity in commerce
41 units
Distribution
Nationwide Distribution: States of IN and TX
Product codes and lots
Part number 262261, Lot number 1925391