Z-2381-2010

Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively

Abbott Medical / initiated 2010-03-26 / Terminated / root cause: design

Hospira Inc began a recall of Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively on . The reason FDA records is "If the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2381-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively.
Product code
FRN Pump, Infusion
Reason for recall
If the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result.
Root cause tag
design
FDA root cause
Device Design
Action
Hospitra Urgent Device Correction Notification mailed on 03/26/2010 via federal express. Tracking done by delivery tracking. Distribution of product continues and has not stopped. Additional notification was issued in 09/2010. A software correction is planned.
Quantity in commerce
28485 units
Distribution
Worldwide distribution: USA, Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.
Product codes and lots
List number configurations: 16026-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16026-27-89/90, 16026-13-25/26, 53/54; 16027-27-89/90; 16027-13-25/26, 53/54.