Z-2285-2010

Suros ATEC 0912-20 Breast Biopsy and Excision System Handpiece, REF ATEC0912-12, sterile, Hologic, Inc. 6100...

Hologic, Inc. / initiated 2010-08-11 / Terminated / root cause: packaging

Hologic, Inc. began a recall of Suros ATEC 0912-20 Breast Biopsy and Excision System Handpiece, REF ATEC0912-12, sterile, Hologic, Inc. 6100 Technology Center Drive Indianapolis, IN. The... on . The reason FDA records is "The firm received a customer complaint that the a device labeled as ATEC 0912-12 was actually ATEC 0912-20". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2285-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hologic, Inc.
Product
Suros ATEC 0912-20 Breast Biopsy and Excision System Handpiece, REF ATEC0912-12, sterile, Hologic, Inc. 6100 Technology Center Drive Indianapolis, IN. The handpiece is inserted into the breast to the site of interest. When the proper location/position is acheived, the handpiece cuts and removes biopsy tissue.
Product code
KNW Instrument, Biopsy
Reason for recall
The firm received a customer complaint that the a device labeled as ATEC 0912-12 was actually ATEC 0912-20.
Root cause tag
packaging
FDA root cause
Packaging
Action
The firm, Hologic, Inc., sent out "Urgent Medical Device Recall" letters by FedEx on August 11, 2010, to all customers. The letter describes the product, problem and action to be taken by the customers. The customers were instructed to check their inventories and to quarantine any of the recalled product; if the product was distributed further, identify the users and notify them at once of this product recall; segregate all products from this lot and quarantine them in preparation for return to Hologic, and complete and return the enclosed RECALL RESPONSE FORM by fax at 877-574-3255 or mail. Note: Replacement product cannot be sent until Hologic receives your response form. If you have any questions, please contact Hologic directly at 317-344-7630.
Quantity in commerce
990
Distribution
Worldwide distribution: USA and countries including Canada, Belgium, Algeria, Germany, London, Lebanon and Italy.
Product codes and lots
00316.