Z-2516-2010
All Med Medical 20ga x .75" Winged Administration Set w/Y Site Model # M2075Y Reorder # 50-2022
Multi-Med, Inc. / initiated 2010-08-30 / Terminated / root cause: design
Multi-Med, Inc. began a recall of All Med Medical 20ga x .75" Winged Administration Set w/Y Site Model # M2075Y Reorder # 50-2022 on . The reason FDA records is "Huber needles may core and result in port leakage or emboli being flushed our of the port". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2516-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Multi-Med, Inc.
- Product
- All Med Medical 20ga x .75" Winged Administration Set w/Y Site Model # M2075Y Reorder # 50-2022
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- Huber needles may core and result in port leakage or emboli being flushed our of the port.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Multi-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
- Quantity in commerce
- 45,537 units
- Distribution
- Worldwide Distribution -- USA and Mexico.
- Product codes and lots
- Lot codes: 5J012A 6A003A 6B056A 6E004A 6E011A 6H052A 6K026A 6L005A 7A016A 7B060A 7C027A 7D001A 7F008A 7I010A 7L026A 8J065A 9E024A 9F020A 9G033A 9H018A