Z-2519-2010
All Med Medical 20ga x 1" Winged Administration Set w/Y Site Model # M201Y Reorder # 50-2032
Multi-Med, Inc. / initiated 2010-08-30 / Terminated / root cause: design
Multi-Med, Inc. began a recall of All Med Medical 20ga x 1" Winged Administration Set w/Y Site Model # M201Y Reorder # 50-2032 on . The reason FDA records is "Huber needles may core and result in port leakage or emboli being flushed our of the port". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2519-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Multi-Med, Inc.
- Product
- All Med Medical 20ga x 1" Winged Administration Set w/Y Site Model # M201Y Reorder # 50-2032
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- Huber needles may core and result in port leakage or emboli being flushed our of the port.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Multi-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
- Quantity in commerce
- 31,261 units
- Distribution
- Worldwide Distribution -- USA and Mexico.
- Product codes and lots
- Lot code: 5G055A 5J034A 5J036A 6H027A 6I010A 6J053A 6K027A 7J008A 8D025A 8F029A 8L043A 9A024A