Z-2533-2010

All Med Medical 22ga x 1.5" Standard Bore Winged Administration Set Model #M2215S Reorder #50-2243

Multi-Med, Inc. / initiated 2010-08-30 / Terminated / root cause: design

Multi-Med, Inc. began a recall of All Med Medical 22ga x 1.5" Standard Bore Winged Administration Set Model #M2215S Reorder #50-2243 on . The reason FDA records is "Huber needles may core and result in port leakage or emboli being flushed our of the port". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2533-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Multi-Med, Inc.
Product
All Med Medical 22ga x 1.5" Standard Bore Winged Administration Set Model #M2215S Reorder #50-2243
Product code
FMI Needle, Hypodermic, Single Lumen
Reason for recall
Huber needles may core and result in port leakage or emboli being flushed our of the port.
Root cause tag
design
FDA root cause
Device Design
Action
Multi-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
Quantity in commerce
672 units
Distribution
Worldwide Distribution -- USA and Mexico.
Product codes and lots
Lot code: 7D061A 7G037A