Z-2657-2010
AVI Sterile Impad Rigid Sole Foot Cover, Left Foot
Medtronic, Inc. / initiated 2010-08-11 / Terminated / root cause: sterility
Covidien LP began a recall of AVI Sterile Impad Rigid Sole Foot Cover, Left Foot Product Code:6067 on . The reason FDA records is "Covidien issued an Urgent Product Recall of the Sterile ImPad Rigid Sole Foot Cover. The product is indicated for prevention of blood pooling in the limb. 120 cases of each Lot code 6066 (right foot) and 6067 (left...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-2657-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- AVI Sterile Impad Rigid Sole Foot Cover, Left Foot Product Code:6067
- Product code
- JOW Sleeve, Limb, Compressible
- Reason for recall
- Covidien issued an Urgent Product Recall of the Sterile ImPad Rigid Sole Foot Cover. The product is indicated for prevention of blood pooling in the limb. 120 cases of each Lot code 6066 (right foot) and 6067 (left foot) were shipped to the distributor before sterilization. The customer is instructed to quarantine the product and return it to the manufacturer for replacement. If there are any
- Root cause tag
- sterility
- FDA root cause
- Release of Material/Component prior to receiving test results
- Action
- Covidien issued an Urgent Product Recall of the Sterile ImPad Rigid Sole Foot Cover. The product is indicated for prevention of blood pooling in the limb. 120 cases of each Lot code 6066 (right foot) and 6067 (left foot) were shipped to the distributor before sterilization. The customer is instructed to quarantine the product and return it to the manufacturer for replacement. If there are any questions please contact, Covidien Customer Service at (800) 962-9888, Option 1.
- Quantity in commerce
- 1436 units (359 cases)
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 010/09, 011/09, 012/09, 013/09