Z-2657-2010

AVI Sterile Impad Rigid Sole Foot Cover, Left Foot

Medtronic, Inc. / initiated 2010-08-11 / Terminated / root cause: sterility

Covidien LP began a recall of AVI Sterile Impad Rigid Sole Foot Cover, Left Foot Product Code:6067 on . The reason FDA records is "Covidien issued an Urgent Product Recall of the Sterile ImPad Rigid Sole Foot Cover. The product is indicated for prevention of blood pooling in the limb. 120 cases of each Lot code 6066 (right foot) and 6067 (left...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-2657-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
AVI Sterile Impad Rigid Sole Foot Cover, Left Foot Product Code:6067
Product code
JOW Sleeve, Limb, Compressible
Reason for recall
Covidien issued an Urgent Product Recall of the Sterile ImPad Rigid Sole Foot Cover. The product is indicated for prevention of blood pooling in the limb. 120 cases of each Lot code 6066 (right foot) and 6067 (left foot) were shipped to the distributor before sterilization. The customer is instructed to quarantine the product and return it to the manufacturer for replacement. If there are any
Root cause tag
sterility
FDA root cause
Release of Material/Component prior to receiving test results
Action
Covidien issued an Urgent Product Recall of the Sterile ImPad Rigid Sole Foot Cover. The product is indicated for prevention of blood pooling in the limb. 120 cases of each Lot code 6066 (right foot) and 6067 (left foot) were shipped to the distributor before sterilization. The customer is instructed to quarantine the product and return it to the manufacturer for replacement. If there are any questions please contact, Covidien Customer Service at (800) 962-9888, Option 1.
Quantity in commerce
1436 units (359 cases)
Distribution
Nationwide
Product codes and lots
Lot Numbers: 010/09, 011/09, 012/09, 013/09