Z-2557-2010
Marquette Medical Huber Trap-A Safety Huber Needle 20ga x .75" Winged Administration Set Model # MW203...
Multi-Med, Inc. / initiated 2010-08-30 / Terminated / root cause: design
Multi-Med, Inc. began a recall of Marquette Medical Huber Trap-A Safety Huber Needle 20ga x .75" Winged Administration Set Model # MW203 Intended to be used to inject fluids into, withdraw... on . The reason FDA records is "Huber needles may core and result in port leakage or emboli being flushed our of the port". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2557-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Multi-Med, Inc.
- Product
- Marquette Medical Huber Trap-A Safety Huber Needle 20ga x .75" Winged Administration Set Model # MW203 Intended to be used to inject fluids into, withdraw fluids from inside the vascular space through a subcutaneously implanted port.
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- Huber needles may core and result in port leakage or emboli being flushed our of the port.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Multi-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
- Quantity in commerce
- 14,775 units
- Distribution
- Worldwide Distribution -- USA and Mexico.
- Product codes and lots
- Lot Code: 5I032M 5K034M 6B044M 6C035 6C048M 6C065M 6H004M 7B022M