Z-0059-2011

Closed knot pusher REF 902813, Biomet Sports Medicine, Warsaw, IN. The device is intended to push knots down...

Zimmer GmbH / initiated 2010-08-02 / Terminated / root cause: design

Biomet, Inc. began a recall of Closed knot pusher REF 902813, Biomet Sports Medicine, Warsaw, IN. The device is intended to push knots down a limb of suture on . The reason FDA records is "The firm has received four complaints that the closed knot pusher is missing a radius which contributes to sutures being cut during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0059-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Closed knot pusher REF 902813, Biomet Sports Medicine, Warsaw, IN. The device is intended to push knots down a limb of suture.
Product code
HXO Pusher, Socket
Reason for recall
The firm has received four complaints that the closed knot pusher is missing a radius which contributes to sutures being cut during surgery.
Root cause tag
design
FDA root cause
Device Design
Action
Biomet Sports Medicine sent URGENT MEDICAL DEVICE RECALL NOTICES dated 8/2/2010 to its customers and distributors. The notices stated that the use of the device should be discontinued and the products returned to Biomet. Healthcare professionals and consumers were to report serious adverse events (side effects) or product quality problems with the use of this product to the FDA's MedWatch Adverse Event Reporting program either online at www.fda.gov/MedWatch/report.htm, by regular mail to MedWatch, 5600 Fishers Lane, Rockville, MD 20852-9787 or by phone at (800)FDA-1088 or Fax (800)FDA-0178. Questions related to this notice should be directed to 574-371-3755 or 574-371-3756, Monday through Friday, 8AM to 5PM.
Quantity in commerce
101
Distribution
Worldwide Distribution - USA, including FL, NY, VA, CA, GA, AL, PA, MI, LA, KS, SD, KY, TX, AR, and WV, and the countries of Belgium, Italy, Argentina, Poland, Netherlands, Turkey and Portugal.
Product codes and lots
098750, 278290, 538080, 666270, 864100 and 948870.