Modular Microplasty Cup inserter 3/8" thread , REF 31-400600
Biomet, Inc. began a recall of Modular Microplasty Cup inserter 3/8" thread , REF 31-400600 on . The reason FDA records is "There was a clerical error in order entry which resulted in the recalling firm recieving spring components from a supplier made of "music wire" The recalling frim conducted tests to determine the effects of theis change...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2504-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Modular Microplasty Cup inserter 3/8" thread , REF 31-400600
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- There was a clerical error in order entry which resulted in the recalling firm recieving spring components from a supplier made of "music wire" The recalling frim conducted tests to determine the effects of theis change in materials. The testing showed that the wire would oxidize and there would be early failure of the device.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm send URGENT MEDICAL DEVICE RECALL NOTICES to its distributors on 7/28/2010. The notices instructed the distributors to discontinue use of the product and return it to Biomet.
- Quantity in commerce
- 331 total for both products
- Distribution
- US, Australia, Canada, Mexico, Chile, Japan.
- Product codes and lots
- 658180, 658190 and 907630.