Z-2505-2010

Modular Microplasty Cup impactor 1/4" - 28 thread insert , REF 31-400605

Zimmer GmbH / initiated 2010-07-28 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Modular Microplasty Cup impactor 1/4" - 28 thread insert , REF 31-400605 on . The reason FDA records is "There was a clerical error in order entry which resulted in the recalling firm recieving spring components from a supplier made of "music wire" The recalling frim conducted tests to determine the effects of theis change...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2505-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Modular Microplasty Cup impactor 1/4" - 28 thread insert , REF 31-400605
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
There was a clerical error in order entry which resulted in the recalling firm recieving spring components from a supplier made of "music wire" The recalling frim conducted tests to determine the effects of theis change in materials. The testing showed that the wire would oxidize and there would be early failure of the device.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The firm send URGENT MEDICAL DEVICE RECALL NOTICES to its distributors on 7/28/2010. The notices instructed the distributors to discontinue use of the product and return it to Biomet.
Quantity in commerce
331 total for both products
Distribution
US, Australia, Canada, Mexico, Chile, Japan.
Product codes and lots
106623, 106624, 106625, 106626, 106627, 106628 110405 and 110423.