Z-0018-2011
Oxylog 3000, Emergency and Transport Ventilator; Catalog #2M86300; Drager Medical Inc., 3135 Quarry Road...
Draeger Medical Systems, Inc. / initiated 2010-08-19 / Terminated / root cause: labeling and UDI
Draeger Medical Systems, Inc. began a recall of Oxylog 3000, Emergency and Transport Ventilator; Catalog #2M86300; Drager Medical Inc., 3135 Quarry Road, Telford, PA 18969 on . The reason FDA records is "Patients received insufficient ventilation. The Oxylog 3000 Instructions are inadequate relating to dead space volume". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0018-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger Medical Systems, Inc.
- Product
- Oxylog 3000, Emergency and Transport Ventilator; Catalog #2M86300; Drager Medical Inc., 3135 Quarry Road, Telford, PA 18969.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Patients received insufficient ventilation. The Oxylog 3000 Instructions are inadequate relating to dead space volume.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- The firm sent Important Safety Notice Letter, dated August 2010, and Instructions for Use supplement to consignees. The letter explained the issue with the affected product. The letter asks customers to attach the Instructions for Use amendment sheet supplied with the letter to their corresponding Instructions for Use. The firm also states that a they will provide a dedicated pediatric breathing circuit for the Oxylog 3000. Questions regarding the letter should be directed to Michael Kelhart at 1-800-543-5047. Questions regarding the use of the device should be directed to DragerService Technical Support at 1-800-543-5047 or your local Drager Sales Representative.
- Quantity in commerce
- 78 potentially afftected units
- Distribution
- Nationwide Distribution
- Product codes and lots
- only units manufactured before 9/08