Z-2462-2010

GCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX...

Abbott Medical / initiated 2010-08-20 / Terminated / root cause: design

Abbott Point of Care Inc. began a recall of GCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX Mountable Downloader-Recharger (GCX) is a... on . The reason FDA records is "Reliability issues associated with broken connection pins within the downloader that mate with the pins from the analyzer". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2462-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
GCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX Mountable Downloader-Recharger (GCX) is a connectivity component that is intended to be mounted to a GCX compatible mount or "arm" and provide a data connection to an authorized manufacturer's patient monitoring systems.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
Reliability issues associated with broken connection pins within the downloader that mate with the pins from the analyzer.
Root cause tag
design
FDA root cause
Device Design
Action
Abbott Point of Care, Inc. sent Urgent Recall Notices, dated August 20, 2010, by Federal Express to two Clarion Health locations. The letter identified the product, the issue, and the actions to be taken by the firm and the customers. 1) An Abbott Point of Care representative will contact the customers to plan the replacement of affected product at their facility. 2) Replacement will occur in batches of approximately 50 units at a time. 3) Abbott will provide personnel to perform the replacements to help expedite the replacement. 4) The customer is required to acknowledge that all units have been replaced by signing the form provided with the Urgent Recall Notice.. If you have any questions regarding this information please contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or your Abbott Point of Care representative.
Quantity in commerce
277
Distribution
Nationwide Distribution - USA, including the state of Indiana
Product codes and lots
List number: 06F23-57/514005, part number 016060-01.