Z-0060-2011

Zimmer Trabecular metal Modular Acetabular system 56 mm Porous shell with cluster holes, Catalog...

Zimmer GmbH / initiated 2010-08-18 / Terminated

Zimmer Inc. began a recall of Zimmer Trabecular metal Modular Acetabular system 56 mm Porous shell with cluster holes, Catalog 00-6202-056-22, Zimmer Inc., Warsaw, IN. This device is... on . The reason FDA records is "The threads on the device were not completely machined and therefore would not mate with athe corresponding cup positioner/shell inserter". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0060-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Trabecular metal Modular Acetabular system 56 mm Porous shell with cluster holes, Catalog 00-6202-056-22, Zimmer Inc., Warsaw, IN. This device is indicated for primary or revision surgery for rehabilitating hips damaged as the result of noninflammatory degenerative joint disease, including osteoarthritis, fused hip, avascular necrosis, traumatic arthritis and fracture of the pelvis.
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
The threads on the device were not completely machined and therefore would not mate with athe corresponding cup positioner/shell inserter.
FDA root cause
Other
Action
The firm, Zimmer, made telephone calls using a phone script document entitled "Urgent: Medical Device Recall" Phone Script Call Communication, to the customers that received this product starting on 8/18/2010. The telephone calls were to determine the status of the recalled product. A Zimmer Field Action Strategy notification dated August 9, 2010 was also provided. The notifications/phone script described the product, problem and actions to be taken by the customers. The customers were instructed to remove the 00-6202-056-22, lot 61091653 product from their inventory and return to Zimmer Product Service Department, Attn: Product Service, 1777 West Center Street, Warsaw, IN 45680. Note: If the product had been used, no action was necessary. The returned product will be evaluated to determine how many of the devices had the defect. If you have any questions, please call (574) 371-8852.
Quantity in commerce
24
Distribution
Worldwide distribution:USA including states of: MI, IN, WI, MO, CA, VA, MD, TX, NC, PA, OR, WA, FL, GA, NY, ME, AZ, and UT and country including Canada.
Product codes and lots
61091653