Z-0137-2011

ACL TOP 700 LAS PN 000280030 Bench top, fully automated, random access analyzer designed specifically for in...

Instrumentation Laboratory Co. / initiated 2010-08-13 / Terminated / root cause: software

Instrumentation Laboratory Co. began a recall of ACL TOP 700 LAS PN 000280030 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis... on . The reason FDA records is "Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0137-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Instrumentation Laboratory Co.
Product
ACL TOP 700 LAS PN 000280030 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.
Product code
GKP Instrument, Coagulation, Automated
Reason for recall
Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves
Root cause tag
software
FDA root cause
Software design
Action
Instrumentaion Laboratory notified the US Distributor/Canada via BCI Service Modification Notice (#10793) on August 13, 2010. The Mandatory software upgrade to Verison V4.3.0 in the United States and Canada will be conducted by BCI service representatives and tracked through the distributor, Beckman Coulter (BCI). International Affiliates/Dealers: Worldwide Regulatory Notices were e-mailed to IL international affiliates and dealers, notifying them of the mandatory software upgrade, providing issue description and a risk summary and the support. The upgrades were to be handled through the local representatives.
Quantity in commerce
5 units
Distribution
Worldwide Distribution, USA and Canada
Product codes and lots
Serial Number range: 08120100-10020126