The GMD UNIVERSAL URINARY INCONTINENCE SLING (Code 1010) is supplied sterile and is intended for single...

Generic Medical Devices, Inc. / initiated 2011-02-14 / Terminated / root cause: design

Generic Medical Devices began a recall of The GMD UNIVERSAL URINARY INCONTINENCE SLING (Code 1010) is supplied sterile and is intended for single patient use only. The GMD Universal Sling is a sterile... on . The reason FDA records is "Protective sheath sticks to itself and sleeve breakage has resulted in complaints and MDRs". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1390-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Generic Medical Devices, Inc.
Product
The GMD UNIVERSAL URINARY INCONTINENCE SLING (Code 1010) is supplied sterile and is intended for single patient use only. The GMD Universal Sling is a sterile, single use device for the treatment of female stress urinary incontinence. It is comprised of a polypropylene sheath with a disposable universal low density polyethylene sleeve at each end of attachment of the sling to GMD's single use or reusable trocars. The universal sleeve has three trocar insertion points, the distal and proximal trocar insertion points are for inside-out, bottom-up approaches and a sleeve end trocar insertion point is for outside-in, top-down approaches. The design allows the physician to utilize the device for placement of a polypropylene mesh at the urethra or bladder neck.
Product code
FTL Mesh, Surgical, Polymeric
Reason for recall
Protective sheath sticks to itself and sleeve breakage has resulted in complaints and MDRs.
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm, GMD, issued an "URGENT MEDICAL DEVICE RECALL" letter dated February 14, 2011 to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed not to use any GMD Model 1010 Universal Slings remaining in their inventory; return any Model 1010 slings in their possession to the Distributors or GMD; notify their customers to whom they shipped Model 1010s of this voluntary recall and request them to return any Model 1010 slings in their possession to them; and return all Model 1010 inventories to GMD using the attached Recalled Model 1010 Return Form. Note: Full credit in the amount of the returned products purchase price will be issued for all sterile, unexpired product returned to GMD. Contact the GMD Product Performance Senior Manager, at phone: (253) 853 3512 or email: [email protected], if you have any questions/issues related to this recall.
Quantity in commerce
approximately 1711
Distribution
Worldwide distribution: USA and countries including: Australia, Argentina, Brazil, Canada, South Africa, and Scandinavia.
Product codes and lots
All model number 1010 Universal Sleeved Slings. It cannot be determined whether each Sling is identified by an individual serial number.