LIBERTY Automated Peritoneal Dialysis System lntegrated Cycler Set with single stay safe patient connector...
Fresenius Medical Care Holdings, Inc. began a recall of LIBERTY Automated Peritoneal Dialysis System lntegrated Cycler Set with single stay safe patient connector and extended drain line PIN: 050-87216. manufactured... on . The reason FDA records is "Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0105-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- LIBERTY Automated Peritoneal Dialysis System lntegrated Cycler Set with single stay safe patient connector and extended drain line PIN: 050-87216. manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, Mexico.
- Product code
- FKX System, Peritoneal, Automatic Delivery
- Reason for recall
- Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Fresenius Medical Care issued an Urgent FMCNA LIBERTY Cycler Set Recall letter on September 3, 2010 identifying the specific affected part numbers and product lots and requesting that customers identify and discontinue the use of the affected lots immediately. Customers were requested to return a receipt of notification. Customers may arrange return unused product to FMCNA by contacting customer service at 1-800-323-5188.
- Quantity in commerce
- 5799 units
- Distribution
- Nationwide Distribution: Throughout USA.
- Product codes and lots
- Lot Number Expiration Date 09ER08101 05/31/2012 09HR08002 06/30/2012 09HR08003 06/30/2012 09JR08057 07/31/2012 09JR08058 07/31/2012 09JR08059 07/31/2012 09KR08002 08/31/2012 09KR08003 08/31/2012 09KR08004 08/31/2012 09KR08005 08/31/2012 09LR08004 09/30/2012 09LR08005 09/30/2012 09LR08006 09/30/2012 09LR08007 09/30/2012 09LR08008 09/30/2012 09LR08010 09/30/2012 09LR08105 09/30/2012