Z-0054-2011
Scorpio/ NRG 4 i n1 Cutting Block The Scorpio total knee system is indicated for primary Total knee...
Howmedica Osteonics Corp / initiated 2010-08-23 / Terminated / root cause: design
Stryker Howmedica Osteonics Corp. began a recall of Scorpio/ NRG 4 i n1 Cutting Block The Scorpio total knee system is indicated for primary Total knee Arthoplasty (TKA) procedures. The Scorpio Femoral Component... on . The reason FDA records is "Stryker has received reports in which the Cutting Blocks have fractured on impaction during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0054-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Scorpio/ NRG 4 i n1 Cutting Block The Scorpio total knee system is indicated for primary Total knee Arthoplasty (TKA) procedures. The Scorpio Femoral Component is implanted on femur after making bone cuts to accommodate the shape of the implant. The Scorpio instrumentation includes a 4 in 1 Cutting block 98010-00xx) which serves as a guide to allow the surgeon to make the bone e cuts necessary for seating the femoral component (implant) in place. The Scorpio Cutting blocks (8010-00xx) are designed to align with two previously located holes and for seating on the distal surg ace of the femur, and provide guided slots for the sawblades to make bone cuts.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Stryker has received reports in which the Cutting Blocks have fractured on impaction during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Stryker Orthopaedics sent an "URGENT PRODUCT RECALL" letter dated August 30, 2010, to all consignees/customers (branches/agencies, hospital risk management, hospital Chief of Orthopaedics, surgery) including foreign countries. The letter described the product, problem, and action to be taken by customers. The customers were instructed to complete and return the attached PRODUCT RECALL ACKNOWLEDGMENT FORM via fax to 201-831-6069. If you have any questions, feel free to contact the Manager, Divisional Regulatory Compliance (201) 972-2100.
- Quantity in commerce
- 839 (total)
- Distribution
- Worldwide distribution: USA and countries including: Asia, Canada, Europe, Middle East and Africa
- Product codes and lots