Z-0139-2011
Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista...
Siemens Healthcare Diagnostics Inc. / initiated 2010-08-19 / Terminated
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System on . The reason FDA records is "High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0139-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.
- Product code
- CEM Electrode, Ion Specific, Potassium
- Reason for recall
- High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System.
- FDA root cause
- Other
- Action
- Siemens sent an Urgent Field Safety Notice (dated "August 2010") to customers identifying the product, the problem and the action to be taken. Customers were instructed on how to continue testing until new lots of the recalled product were available and asked to complete and fax a Field Correction Effectiveness Check to 301-631-8467 to confirm receipt of the Urgent Field Safety Notice. For technical questions or concerns customers were instructed to contact the Siemens Technical Solutions Center at 800-441-9250.
- Quantity in commerce
- 3,706 cartons
- Distribution
- USA, Nationwide Distribution
- Product codes and lots
- KA0C03 EXP 3/24/11, KA0D01 EXP 4/14/11, KA0E01 EXP 5/5/11, KA0E02 EXP 5/19/11, KA0F01 EXP 6/3/11, AND KA0G01 EXP 7/8/11