Z-0021-2011

The BABIX holder is an accessory used with the BuckyDiagnost X-ray system. It is used to hold the baby either...

Philips North America LLC / initiated 2010-09-09 / Terminated / root cause: manufacturing process

Philips Medical Systems North America Co. Phillips began a recall of The BABIX holder is an accessory used with the BuckyDiagnost X-ray system. It is used to hold the baby either in a secured or immobilized standing position in... on . The reason FDA records is "Potential of broken strands in the main steel cable supporting the accessory holder used with BuckyDiagnost and DigitalDiagnost X-ray systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0021-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
The BABIX holder is an accessory used with the BuckyDiagnost X-ray system. It is used to hold the baby either in a secured or immobilized standing position in front of the X-ray detector during the examination.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Potential of broken strands in the main steel cable supporting the accessory holder used with BuckyDiagnost and DigitalDiagnost X-ray systems.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Philips issued URGENT-Device Correction Notice letters dated September 9, 2010 to their consignees identifying the affected products and action to be taken by customers. The letter states that Babix holder used with Philips BuckyDiagnost and DigitalDiagnost X-Ray System - VS Wall Stand Risk of steel cable fracture in the support holder. The consignees are informed that there is a potential of broken strands at the main steel cable within the holder which cannot be recognized by the user. The firm will issue the Field Change Order (FCO) and the Service Engineer will replace the Babix holder. Consignees can call Philips Call Center at 800-722-9377, #5 (then follow prompts) and reference to FCO 70400038 for the BuckyDiagnost X-ray system and FCO 71200046 for the DigitalDiagnost X-ray system.
Quantity in commerce
4 units
Distribution
Nationwide Distribution in the states of CA, GA, and TN.
Product codes and lots
Devices are identified as Site Numbers: 546607, 48682949, 554660, and 555738.