Z-0627-2011
Wingspan Stent System, Model Number M003WE0300150, Manufactured by Boston Scientific, Fremont, CA Indicated...
Boston Scientific Corporation / initiated 2010-08-25 / Terminated / root cause: manufacturing process
Boston Scientific Corporation began a recall of Wingspan Stent System, Model Number M003WE0300150, Manufactured by Boston Scientific, Fremont, CA Indicated for use in improving cerebral artery lumen diameter... on . The reason FDA records is "Products may not meet radial force specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0627-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Wingspan Stent System, Model Number M003WE0300150, Manufactured by Boston Scientific, Fremont, CA Indicated for use in improving cerebral artery lumen diameter in patients with intracranial atherosclerotic disease, refractory to medical therapy, in intracranial vessels with less than or equal to 50% stenosis that are accessible to the system.
- Product code
- NJE Intracranial Neurovascular Stent
- Reason for recall
- Products may not meet radial force specifications.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Recall initiated August 25, 2010. "Urgent Medical Device Recall - Immediate Action Required" notifications, dated 25 August 2010, were delivered by FedEx where possible, and by personal delivery assigned to sales representatives. The letter identified the affected product and the reason for recall. The letter stated that further distribution or use of any remaining product should cease immediately. Customers were asked to read the enclosed instructions and told that their local Sales Rep could answer any questions that they may have regarding the recall.
- Quantity in commerce
- 10 units
- Distribution
- International Distribution Only -- China, Germany, Italy, the Philippines, Portugal and Turkey.
- Product codes and lots
- Batch numbers 13215033 and 13192372, Catalog/product UPN number M003WE0300150